Hygiena SystemSURE Plus (ATP Meter Only)

Regular price $4,449.91

SKU:HGSS3

Hygiena's™ SystemSURE Plus is the world's best-selling ATP Sanitation Monitoring System (Hygiena's description). With improved functionality, new software, and increased memory, the SystemSURE Plus meets the demands of the cleaning and restoration companies around the world.

This listing is the meter only (Hygiena code SS3). If you are starting from nothing, the SystemSURE Plus Starter Kit adds 100 UltraSnap swabs, a carrying case and a meter stand.

What Is ATP Testing?

The ATP test is a process of rapidly measuring actively growing microorganisms through the detection of adenosine triphosphate, or ATP.

ATP is a molecule found in and around living cells, and as such it gives a direct measure of biological concentration and health. ATP is quantified by measuring the light produced through its reaction with the naturally occurring firefly enzyme luciferase using a luminometer. The amount of light produced is directly proportional to the amount of ATP present in the sample.

What a reading is and is not: ATP is present in organic residue generally — food soil, body fluids, biofilm and the cells within them — not only in living microorganisms. A reading in relative light units (RLU) tells you how much organic residue is on the surface. It does not identify organisms and it is not a microbial count. Use it to verify that cleaning removed the soil; use laboratory microbiological sampling where a job needs organisms identified or counted.

Why Use ATP Testing?

With ATP testing, a restoration contractor can test the initial condition of surfaces and standing water to determine the level of contamination and thus categorize the loss more accurately as 1, 2 or 3.

The testing can also be used during the cleaning and drying processes to determine progress. And finally, at the completion of the job, testing can verify clean and sanitary materials and surfaces.

Keep the categorisation point in proportion: an ATP result is supporting evidence, not the classification itself. Water category is decided from the source of the water, how long it has been there and what it has contacted, as your restoration standard sets out. A high reading supports treating the loss as contaminated; a low reading on its own does not make a loss Category 1.

SystemSURE Plus Benefits

As listed by Hygiena:

  • Most cost-effective ATP hygiene monitoring system available
  • Provides real-time feedback and facilitates long-term data analysis
  • Optimizes and verifies that cleaning procedures are working
  • Assists in developing and improving process and risk assessment programs (ie. HACCP & GMPs)
  • No scheduled yearly maintenance is necessary
  • Available Calibration Control Kit for in-house confirmation of calibration
  • Complimentary SureTrend Software

SystemSURE Plus Specifications

  • Sensitive - can detect down to 1 femtomole (1 x 10-15 moles) of ATP
  • Advanced photodiode sensor offers low variation and stays in calibration
  • Calibration self-check at start up with optional full diagnostic check mode
  • The symbol keypad is easy for any user to understand and navigate
  • Powered by 2 x AA batteries for several months of uninterrupted use
  • Memory capacity:
    • 100 programmable test plans
    • 200 programmable user IDs
    • 251 programmable locations per test plan
    • 5,000 programmable test locations
  • Small, lightweight, handheld instrument (0.57 lbs., 7.6 x 17.8 x 3 cm)
  • Quick-test mode enables ad-hoc testing
  • Large viewable screen with back-light and optional bold text
  • Removable read chamber for easy cleaning
  • Connects to SureTrend software via USB

At a Glance

Brand / model Hygiena SystemSURE Plus (SS3)
SKU HGSS3
Sensor Photodiode
Detection limit 1 femtomole (1 × 10-15 mol) ATP
Result time About 15 seconds (Hygiena)
Power 2 × AA batteries — Hygiena quotes up to 8–12 months of use
Size 7.6 × 17.8 × 3 cm
Weight 0.57 lb (about 260 g)
Results memory 2,000 test results (Hygiena)
Data connection USB to SureTrend software (cable sold separately)
Calibration check Start-up self-check; CalCheck for in-house verification (sold separately)
Warranty Not published by Hygiena — ask CCW

Which Tests It Reads

Hygiena lists the SystemSURE Plus as compatible with its UltraSnap surface ATP test and AquaSnap water ATP test, and with its other luminometer-read test devices. The Hygiena tests CCW stocks:

The PRO-Clean and SpotCheck Plus tests are colour-change tests read by eye — they do not need the meter.

Third-party swabs: Anaeron lists its CleanTrust ATP Surface Swab as compatible with the SystemSURE Plus. Hygiena lists only its own tests. If you mix brands, set a separate baseline and limits for each swab type — RLU figures from different swab chemistries are not interchangeable.

Using It Well

  • Let the tests warm up. Hygiena's UltraSnap instructions call for the device to reach 21–25 °C before testing. A swab straight from the fridge reads low.
  • Swab the same way every time. A 10 × 10 cm area in a criss-cross pattern is Hygiena's standard for flat surfaces. Results are only comparable when the sampling is.
  • Read promptly. Insert the activated swab within 30 seconds (UltraSnap instructions). The SystemSURE Plus gives the result in about 15 seconds.
  • Set your own limits. The meter supports programmable pass, caution and fail limits. Build them from readings on surfaces you know are properly cleaned, not from a number copied from someone else's procedure.
  • Sanitiser still on the surface can suppress the reading. Sample after cleaning and before sanitising, or allow for it in your procedure.

Calibration and Data

The meter runs a sensor self-check at start-up. For in-house confirmation between checks, Hygiena's reusable CalCheck replaces the older PCD4000 Calibration Control Kit. A CalCheck verifies the meter; it does not adjust it — a failed check means contact CCW.

Results transfer to Hygiena's SureTrend software over USB using the SystemSURE Plus USB cable. Hygiena now also offers SureTrend as a cloud platform; confirm with CCW which version and licence applies before you build a reporting system around it.

Frequently Asked Questions

What comes with the meter-only listing?

The luminometer. Hygiena supplies the SystemSURE Plus with a protective carrying case and shoulder strap; confirm with CCW what is in the box on current stock. Swabs, the USB cable and the CalCheck are sold separately.

Does it need annual calibration?

Hygiena says no scheduled yearly maintenance is needed. Use the CalCheck to confirm the meter is reading correctly between checks.

Can it tell me whether a surface has bacteria or mould on it?

No. It measures ATP, which is present in organic residue generally. It shows how clean a surface is, not which organisms are on it.

SystemSURE Plus or EnSURE Touch?

Both read Hygiena's ATP tests. The EnSURE Touch has a touch screen, Wi-Fi data transfer and a 10-second read. The SystemSURE Plus is the simpler keypad meter with USB download.

Can I use CleanTrust swabs in it?

Anaeron says yes; Hygiena lists only its own tests. See "Which Tests It Reads" above, and keep separate limits for each swab brand. CleanTrust also make their own Touch YZ3500 luminometer.

Related Products